<< Click to Display Table of Contents > Analog Output And Power Supply |
The 16 bit analog outputs are provided as a differential signal for the X, Y and Z channels via the 10-pin connector SV1. For operation an external +/- 15 V supply must be connected. Configuration of the channels to either +/-5 V or +/- 10 V output range is possible as is the delivery of 15 mA output current.
The negative signals are used for scan heads with differential inputs. When using a scan head without differential inputs the positive signals are used in respect to ground (pin 6 or 10 AGND). When using the differential outputs signals with a scanner without differential inputs the scanning field will enlarge to the double size. |
---|
Figure 6: SV1 - 10-pin analog connector pin assignment
Pin |
Name |
Comments |
---|---|---|
1 |
AV+ (+15 V) |
+15 V positive supply in respect to AGND, 100 mA max. current Appropriate fuse for circuit protection must be provided by the external circuit. |
2 |
+X |
Positive X channel, max. 15 mA current |
3 |
-X |
Negative X channel, max. 15 mA current |
4 |
+Y |
Positive Y channel, max. 15 mA current |
5 |
-Y |
Negative Y channel, max. 15 mA current |
6 |
AGND |
Analog Ground, connected also to pin 10 of SV1 |
7 |
-Z |
Negative Z channel, max. 15 mA current |
8 |
+Z |
Positive Z channel, max. 15 mA current |
9 |
AV- (-15 V) |
-15 V negative supply in respect to AGND, 50 mA max. current Appropriate fuse for circuit protection must be provided by the external circuit. |
10 |
AGND |
Analog Ground, connected also to pin 6 of SV1 |